Features and Benefits
90℃ scattered light method
Reliable concentration measurement
Isolated 4-20mA output, RS485
Air cleaning device (optional)
Simple, user friendly menu
Immediate installation
Applications
Process water
Cycle water
Waste water
Pure water
Drinking water
Technical Data
1.Measurement input: Turbidity/SS, Temperature
2.Measuring range:
Turbidity 0.0-4000NTU Resolution 0.1 to 1NTU
SS 0.0-4500.0mg/L Resolution 0.1 to 1mg/L
3.Temperature Compensation: -10℃-130℃ ATC/MTC; Pt-1000
4.Environmental Conditions: Operating temperature: 0-70℃
Storage temperature:-20℃-70℃
Relative Humidity: 0-90% NC
5.Electrical connection: 90-260V AC, 50/60Hz; 24VDC(Optional)
6.Current output
Turbidity/SS current output1: 4-20mA output,max.load 500Ω
Temperature current output2: 4-20mA output,max.load 500Ω
Current Output accuracy: ±0.05 mA
7.Relay out
Relay1&2(Alarm) Operation: ON/OFF
Selectable action: High/low
Relay3(Washing) Wash interval:0.1-1000 h
Wash time: 1-1000s
8.Interface: RS485 Modbus RTU Baud rate:9600/19200/38400
9.Maximum relay contacts capacity: 5A/250VAC, 5A/30VDC
10.Enclosure rating: IP65
11.Display: Back light, dot matrix, LCD
12.Language: Chinese/English
13.Password Set model:1200
Calibration model:1100
14.Installation type: Wall/Panel/Pipe Mounted
15.Dimensions: 144×144×110 mm
16.Hole Size: 138×138 mm
17.Weight: 0.86 kg